1) Regeneration gas: a mixture of hydrogen and inert gas (hydrogen accounts for five% -10%)

2) Quantity: 40L gasoline cylinder 10Mpa

3) Regenerative gasoline tension: .05Mpa (the gasoline intake per 50 % hour in the condition of .05Mpa is that the cylinder pressure drops by 1.5Mpa. If it exceeds 1.5Mpa, the strain of the tension regulator’s program will be diminished, if it is less than one.5Mpa Enhance the stress of the force regulator schedule)

4) System: three hours heating-3 hours heating and innovative technology glovebox air flow-three several hours vacuuming-1 hour pumping-10 several hours cooling

5) Techniques:

① Link the regeneration gas (verify the pipeline for leaks soon after link: soon after connecting, open the main valve of the strain lessening valve to the highest, open the auxiliary meter to .1Mpa, flip off the major valve, wait around 15 minutes, if the strain of the auxiliary meter stays at .one Mpa remains unchanged, indicating no air leakage

>②Set the stress of the auxiliary gauge of the stress lessening valve to .1Mpa (the tension through air flow is about .06-.08MPa

>③ Near the cycle and analyz

r>④Turn on the vacuum pu

r>⑤ Commence regenerat

br>⑥ The new system finishes at 20h (clean up in advance of cleaning, drinking water and oxygen index <200p


6) Note: Do not cut off the power and gas during the regeneration process (especially the first 6 hours)! Some pressure reducing valves are not allowed, and the pressure needs to be adjusted appropriately to ensure continuous ventilation for 3 ho


7) Observation: whether the regeneration gas flowmeter shows about 20L / min, otherwise adjust the pressure of the pressure reducing valve.